Product Description
Features
.Support logic analyzer and multimeter
.2GSa/s sampling rate and 50GSa/s equivalent sampling rate
.1024k channel memory depth
.5.7-inch color LCD display
.USB-host/device and option LAN, GPIB interface
| Technical Data | DQ1042C | DQ1062C | DQ1102C | DQ1152C | DQ1202C |
| Channels | 2 external + logical analyzer + multimeter | ||||
| Sampling rate | 2GSa/s | ||||
| Equivalent sampling rate | 50GSa/s | ||||
| Display | |||||
| Type | 5.7″ rectangle color LCD | ||||
| Backlight intensity | 300 ni | ||||
| Display resolution | 320 horizontal x 240 vertical pixels | ||||
|
Vertical system |
|||||
| Sensitivity and accuracy | 1mV/div~10V/div | ||||
| Vertical resolution | 8 bit | ||||
| Width of band (-3dB) | DC (AC 5Hz) ~ 40MHz |
DC (AC 5Hz) ~ 60MHz |
DC (AC 5Hz) ~ 100MHz |
DC (AC 5Hz) ~ 150MHz |
DC (AC 5Hz) ~ 200MHz |
| Selectable analog bandwidth limit |
20MHz | ||||
| Rise time | ≤8.7ns | ≤5.8ns | ≤3.5ns | ≤2.3ns | ≤1.8ns |
| DC gain Accuracy | ±5%(1mV/div) | ±4%(2mV/div) | ±3%(5mV/div~10V/div) | ||
| DC measurement accuracy | ±(5%+0.1div)(1mV/div) | ±(5%+0.1div)(1mV/div) | ±(5%+0.1div)(1mV/div) | ||
|
Horizontal system |
|||||
| SEC/DIV range | 2ns~50s/div(200MHz, 150MHz) | ||||
| 5ns~50ns/div(100MHz~40MHz), at 1-2-5 increment | |||||
| Waveform interpolation | Sin(x)/x | ||||
| Recording length | 1024k | ||||
| Sampling rate and delay time accuracy |
±50ppm (any time interval 1ms) | ||||
| Delta time measurement accuracy |
Single: ±(1 sampling time interval + 50ppm rdg. + 0.6ns) | ||||
| Average values: ±(1 sampling time interval + 50ppm rdg. + 0.4ns) | |||||
|
Trigger system |
|||||
| Mode | Auto, normal, single | ||||
| Type | Edge, pulse, video, alternate | ||||
| Hold off range | 100ns~1.5s | ||||
|
Math |
+, -, x, ÷ | ||||
| FFT | |||||
| Acquire input | |||||
| Input coupling | DC, GND, AC | ||||
| Input impedance | 1MΩ±1.5% 24pF±3pF | ||||
| Probe attenuation | 1x, 10x, 100x, 1000x | ||||
| Max. input voltage | 400V (DC+AC peak) | ||||
| Measurement | |||||
| Cursor | Voltage difference ( V) between cursors Time difference ( T) between cursors Reciprocal of T in Hz (1/ T) |
||||
| Auto-measure | Vrms, Vavg, Vp-p, Vmax, Vmin, Vtop, Vhigh, Vlow, Vmid, Vamp; Period, Freq, Rise, Fall, +Width, -Width, +Duty, -Duty, Delay |
||||
| I/O | |||||
| Standard | USB(D), USB(H) | ||||
| Options | GPIB, LAN | ||||
| Calibrator | |||||
| Output voltage | 3V(≥1MΩ load) | ||||
| Output frequency | 1kHz | ||||
| Logical analyzer | |||||
| Input channel | 200MSa/s | ||||
| Max. sampling rate | D0~D15 | ||||
| Recording length | 512k | ||||
| Max. input voltage | ±40Vpeak | ||||
| Minimum voltage slew | 1.2Vpp | ||||
| Trigger mode | Code | D0~D15 Trigger code: H, L, X Clock edge: Rise, Fall |
|||
| Persist | D0~D15 Trigger code: H, L, X Condition: >, <, = Persist setup: 20ns~10s |
||||
|
Digital multimeter characterizes |
|||||
| DC voltage | Range | 400.0mV/4.000V/40.00V/400.0V | |||
| Accuracy | ±(1%+5digits) | ||||
| AC voltage (40Hz~400Hz) | Range | 400.0mV/4.000V/40.00V/400.0V | |||
| Accuracy | ±(1.2%+5digits) | ||||
| Resistance | Range | 400Ω/4kΩ/40kΩ/400kΩ/4MΩ/40MΩ | |||
| Accuracy | ±(1.5%+5digits) | ||||
| On/off | <70.0Ω | ||||
| Diode | Forward voltage drop 0.5~0.8V | ||||
| DC current | Range | 4mA/40mA/400mA | |||
| Accuracy | ±(1%+5digits) | ||||
| Range | 4A | ||||
| Accuracy | ±(1.5%+5digits) | ||||
| Power source | 100~240VACrms, 45~440Hz; 50VA Max; CAT II | ||||
| Dimensions (Wx Hx D) | 320mm x 150mm x 130mm | ||||
| Weight | 2.5kg | ||||
| Accessories | Operation manual, adapter, probex 2, multimeter penx2, current-voltage convertor modulex 2 | ||||